Cleaning the air filter, 3. cleaning the air filter, 5725a – Fluke 5725A User Manual

Page 118: Instruction manual, Figure 6-1. accessing the fuse

Advertising
background image

5725A

Instruction Manual

6-4

100V
110V
115V

120V

S2 S3 S4

VOLTAGE

SELECTION

FUSE

F8A 250V

(FAST)

200V
220V
230V
240V

S2 S3 S4

F4A 250V

(FAST)

JOHN FLUKE MFG. CO., INC.

MADE IN U.S.A.

PATENTS PENDING

NO INTERNAL USER SERVICEABLE
PARTS REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL

TO CLEAN FILTER
REMOVE FROM INSTRUMENT
AND FLUSH WITH WARM
SOAPY WATER

5700A CALIBRATOR

20V PK

MAX

CHASSIS
GROUND

FUSE

WARNING

GROUNDING
CONNECTOR IN POWER CORD
MUST BE CONNECTED TO
ENSURE PROTECTION FROM
ELECTRONIC SHOCK.

47-63 Hz
750VA MAX

CAUTION

FOR FIRE PROTECTION
REPLACE ONLY WITH A 250V FUSE
OF INDICATED RATING.

S4

S3

S2

11A MAX

HI

LO

CALIBRATION

ENABLE NORMAL

CURRENT

OUTPUT

!

!

!

aq13f.eps

Figure 6-1. Accessing the Fuse

6-3. Cleaning the Air Filter

Caution

Damage caused by overheating may occur if the area around
the fan is restricted, the intake air is too warm, or the air filter
becomes clogged.

The air filter must be removed and cleaned every 30 days, or more frequently if the
calibrator is operated in a dusty environment. The air filter is accessible from the rear
panel of the calibrator.

To clean the air filter, refer to Figure 6-2 and proceed as follows:

1.

Turn off the POWER switch and disconnect the line power cord from ac line power.

2.

Squeeze together the sides of the air filter housing and pull it off.

3.

Wash the filter element and housing in soapy water.

4.

Rinse the filter element and housing in fresh running water.

5.

Shake out excess water, then allow the filter element to dry thoroughly before
reinstalling it.

6.

Snap the filter housing back into place.