Dc voltage specifications, 12. dc voltage specifications – Fluke 5725A User Manual

Page 23

Advertising
background image

Introduction and Specifications

DC Voltage Specifications

1

1-7

1-12. DC Voltage Specifications

Table 1-1. 5720A Series II DC Voltage Specifications: 99 % and 95 % Confidence Levels

5720A

99%

99 % Confidence Level

Range

Resolution

Absolute Uncertainty

±

5 °C from calibration temperature

For fields strengths >1 V/m but

3 V/m,

add 0.01% of range.

Relative Uncertainty

±

1 °C

24 Hours

90 Days

180 Days

1 Year

24 Hours

90 Days

±

(ppm output +

µ

V)

±

(ppm output +

µ

V)

220 mV

10 nV

0

5 + 0.5

7 + 0.5

0

8 + 0.5

0

9 + 0.5

0

2 + 0.4

2.5 + 0.4

2.2 V

100 nV

3.5 + 0.8

4 + 0.8

4.5 + 0.8

0

6 + 0.8

0

2 + 0.8

2.5 + 0.8

11 V

1

µ

V

2.5 + 3

3 + 3

3.5 + 3

0

4 + 3

0

1 + 3

1.5 + 3

22 V

1

µ

V

2.5 + 5

3 + 5

3.5 + 5

0

4 + 5

0

1 + 5

1.5 + 5

220 V

10

µ

V

3.5 + 50

4 + 50

0

5 + 50

0

6 + 50

0

2 + 50

2.5 + 50

1100 V

100

µ

V

0

5 + 500

6 + 500

0

7 + 500

0

8 + 500

2.5 + 400

0

3 + 400

5720A

95%

95% Confidence Level

Range

Resolution

Absolute Uncertainty

±

5 °C from calibration temperature

For fields strengths >1 V/m but

3 V/m,

add 0.01 % of range.

Relative Uncertainty

±

1 °C

24 Hours

90 Days

180 Days

1 Year

24 Hours

90 Days

±

(ppm output +

µ

V)

±

(ppm output +

µ

V)

220 mV

10 nV

4 + 0.4

0

6 + 0.4

6.5 + 0.4

7.5 + 0.4

1.6 + 0.4

0

2 + 0.4

2.2 V

100 nV

3 + 0.7

3.5 + 0.7

0

4 + 0.7

0

5 + 0.7

1.6 + 0.7

0

2 + 0.7

11 V

1

µ

V

2 + 2.5

2.5 + 2.5

0

3 + 2.5

3.5 + 2.5

0.8 + 2.5

1.2 + 2.5

22 V

1

µ

V

2 + 4

2.5 + 4

0

3 + 4

3.5 + 4

0.8 + 4

1.2 + 4

220 V

10

µ

V

3 + 40

3.5 + 40

0

4 + 40

0

5 + 40

1.6 + 40

0

2 + 40

1100 V

100

µ

V

4 + 400

4.5 + 400

0

6 + 400

6.5 + 400

0

2 + 400

2.4 + 400

Advertising