Voltage standby, 8. voltage standby, 5725a – Fluke 5725A User Manual

Page 66: Instruction manual, Figure 3-4. voltage standby mode

Advertising
background image

5725A

Instruction Manual

3-10

3-8. Voltage

Standby

NO

OUTPUT

CURRENT

OUTPUT

HI

LO

VOLTAGE MODE

CURRENT MODE

5725A STANDBY

FAULT

OVERLOAD

CABLE OFF

11A

MAX

2OV

PK

MAX

NO OUTPUT

HI

HI

LO

LO

AUX

CURRENT

GUARD

GROUND

WIDEBAND

SENSE

V A

OUTPUT

V A

µ

HI

BOOST

LIT

aq9i.eps

Figure 3-4. Voltage Standby Mode

Voltage standby mode is when both VOLTAGE MODE and 5725A STANDBY
indicators are lit. In voltage standby mode, the 5725A is not supplying an output.
Voltage standby occurs if "5725A" is selected in a 5700A setup menu as the voltage
amplifier, and the next two conditions are true:

An ac voltage 220V or above is showing on the 5700A Output Display with the
STANDBY annunciator lit.

The BOOST indicator on the 5700A is lit.

Note

If condition 1 is true, the BOOST indicator can be toggled on and off by
pressing the 5700A BOOST key. If the BOOST indicator is off, the 5725A is
in standby mode, not voltage standby.