Series 730 Instruction Manual

IM-73-E

9

Problem

Possible Cause

Solution

Flow measurement is
erratic or fluctuating

Very erratic or non-uniform flow

Follow installation requirements
shown in Figure 2

Flow meter installed with less than re-
quired minimum pipe diameters up-
stream and downstream of the sensor

Follow installation requirements
shown in Figure 2

Probe is not secured to a solid base

Secure probe to a non-vibrating solid
mount

Sensor component broken

Return to factory for replacement

Malfunction in system
electronics

Return to factory for evaluation

Moisture present in gas flow

Install a water trap or filter upstream
of the flow sensor

Velocity measurement
seems too high or low

Sensor assembly not aligned correctly
to flow

Correct alignment with the flow indica-
tor pointing downstream in the direc-
tion of flow

No response to flow from
sensor assembly

No power

Turn on power to the flow meter

Flow profile distortions

Try to find another location for the
meter

Extremely turbulent flow

Do not place the meter near a ventila-
tor, static mixer or valve

Sensor failure

Return to factory for repair

Malfunction in system
electronics

Return to factory for evaluation

System will not zero

Out of calibration

Return to factory for calibration

Probe tip is broken

Return to factory for repair

4-20 mA output circuit not
indicating 4 mA at zero
flow

Improper current loop resistance; total
current loop resistance must be no
less than 150 Ohms and no greater
than 450 Ohms

Install load resistor or use larger
gauge wire

Output pegs plus or minus

Flow body not plugged in

Plug in both connectors

Malfunction in system
electronics

Return to factory for evaluation