5-52

System Setup Menu

Setup GPIB

The Setup GPIB key activates the GPIB parameters submenu. GPIB
instrument address and Remote Overide are set in this submenu.

GPIB Address

This key activates the GPIB Address entry field for
numeric and knob entry. The instrument address can be
set from 0 to 30.

Overide Remote ?

In general, every GPIB interface command will put the
SR850 into the REMOTE state with the front panel inac-
tivated. To defeat this feature, set the Overide Remote ?
to Yes. In this mode, the front panel is not locked out
when the unit is in the REMOTE state.

If the SR850 is in the REMOTE state, the [HELP] key
returns the unit to local front panel control.

Return

The Return key will return to the Setup Communications
menu.

Overide Remote?

Return:

No

Yes

GPIB Address

8

Setup RS232

The Setup RS232 key activates the RS232 parameters submenu. Baud
rate, word length, and parity may be configured in this submenu.

Baud Rate

The Baud Rate key allows the knob to adjust the RS232
baud rate. The baud rate can be set to any standard
value from 300 to 19200 baud.

Word Length

This key toggles the character length. The RS232 char-
acter length can be 7 or 8 bits. 8 bits is standard.

Parity

This key toggles the parity. The RS232 parity can be
Even, Odd, or None.

Return

The Return key will return to the Setup Communications
menu.

Return:

7 bits

8 bits

None

Parity

Word Length

Baud Rate

9600 bd

Odd

Even