Remote Programming

OUTPUT and OFFSET COMMANDS

FOUT (?) i {, j}

The FOUT command sets or queries the front panel (CH1 and CH2)
output sources. The parameter i selects CH1 (i=1) or CH2 (i=2) and is
required. The FOUT i, j command sets output i to quantity j where j is
listed below. The definition of j is different for the two outputs.

CH1 (i=1)

CH2 (i=2)

j

output quantity

j

output quantity

0

X

0

Y

1

R

1

R

2

θ

2

θ

3

Trace 1

3

Trace 1

4

Trace 2

4

Trace 2

5

Trace 3

5

Trace 3

6

Trace 4

6

Trace 4

OEXP (?) i {, x, j}

The OEXP command sets or queries the output offsets and expands.
The parameter i selects X (i=1), Y (i=2) or R (i=3) and is required. The
parameter x is the offset in percent (-105.00

x

105.00) and j is the

expand (1

j

256). The OEXP i, x, j command will set the offset and

expand for quantity i. This command requires BOTH x and j. The OEXP?
i command queries the offset and expand of quantity i. The returned
string contains both the offset and expand separated by a comma. For
example, if the OEXP? 2 command returns "50.00,10" then the Y offset
is 50.00% and Y expand is 10.

AOFF i

The AOFF i command automatically offsets X (i=1), Y (i=2) or R (i=3) to
zero. The parameter i is required. This command is equivalent to press-
ing the Auto softkey in the Offset & Expand menu box.