3 performance tests necessary equipment, Warm up, The test record – SRS Labs SR850 User Manual

Page 221: If a test fails

Advertising
background image

7-3

Performance Tests

Necessary Equipment

The following equipment is necessary to complete
the performance tests. The suggested equipment
or its equivalent should be used.

1. Frequency Synthesizer

Freq Range

1 Hz to 1 MHz

Freq Accuracy

better than 5 ppm

Amplitude Accuracy 0.2 dB from 1 Hz to

100 kHz

Harmonic Distortion

-65 dBc

Spurious

-55 dBc

TTL SYNC

available

Recommended

SRS DS345

2. AC Calibrator

Freq Range

10 Hz to 100 kHz

Amplitude

1 mV to 10 V

Accuracy

0.1%

External phase locking capability

Recommended

Fluke 5200A

3. DC Voltmeter

Range

19.999 V, 4 1/2 digits

Accuracy

0.005%

Recommended

Fluke 8840A

4. Feedthrough Terminations

Impedance

50

Warm Up

The lock-in should be turned on and allowed to
warm up for at least an hour before any tests are
performed. The self test does not require any
warm up period.

It is necessary to turn the unit off and on to preset
it. As long as the unit is powered on immediately,
this will not affect the test results.

The Test Record

Make a copy of the SR850 Performance Test
Record at the end of this section. Fill in the results
of the tests on this record. This record will allow
you to determine whether the tests pass or fail and
also to preserve a record of the tests.

If A Test Fails

If a test fails, you should check the settings and
connections of any external equipment and, if pos-
sible, verify its operation using a DVM, scope or
some other piece of test equipment.

After checking the setup, repeat the test from the
beginning to make sure that the test was per-
formed correctly.

If the test continues to fail, contact Stanford
Research Systems for further instructions. Make
sure that you have the unit's serial number and
firmware revision code handy. Have the test
record on hand as well.