6 sr850 service, Adjusting the notch filters – SRS Labs SR850 User Manual

Page 252

Advertising
background image

8-6

SR850 Service

Adjusting the Notch Filters

1. The Notch filter frequency and depth adjustments must be done together. This procedure requires

a BNC cables. A small slotted screwdriver is also required to adjust the potentiometers.

2. Turn the unit ON while holding the [<-] key down. This resets the unit. Let it warm up for at least

an hour.

3. Remove the four black screws which secure the top lid. There are two screws on each side. Slide

the top lid back about 1/4". Remove the lid by pulling the lid towards the back and upwards.

4. The adjustment pots are on the Analog board on right hand side of the card cage. The notch filter

adjustments are accessible through the card cage vent holes. Do not remove the card cage
cover.

5. The adjustment pots are shown in the figure.

The Notch Frequency and Notch Depth adjustment
pots are outside the shield which is mounted on the
analog board.

6. Connect the SINE OUT output to the A input with a BNC cable.

Press the following keys to set the frequency, display and time constant.

[REF/PHASE]

<Ref. Frequency> Select Reference Frequency

[6] [0] [ENTER] Enter 60 Hz OR
[5] [0] [ENTER] Enter 50 Hz

[DISPLAY/SCALE]

<Type/Trace> Select Trace Number for Top display

[3] [ENTER] Enter 3 to display R on the top

<Full/Top/Bottom> to select Bottom display

[4] [ENTER] Enter 4 to display

θ

on the bottom

[GAIN/TC]

<Filter dB/oct> twice to select 24 dB/oct
<Synchronous> to select <200 Hz

After the readings are stable, the top display should read 1.00 V (±2%).

Line Notch Freq

Common Mode

Voltage Input Offset

Current Input Offset

Analog

Board

DSP
Board

2xLine Notch Freq

Line Notch Depth

2xLine Notch Depth

Front