External noise sources – SRS Labs SR850 User Manual

Page 79

Advertising
background image

3-25

SR850 Basics

In addition to the intrinsic noise sources discussed
in the previously, there are a variety of external
noise sources within the laboratory.

Most of these noise sources are asynchronous,
i.e. they are not related to the reference and do
not occur at the reference frequency or its harmon-
ics. Examples include lighting fixtures, motors,
cooling units, radios, computer screens, etc.
These noise sources affect the measurement by
increasing the required dynamic reserve or length-
ening the time constant.

Some noise sources, however, are related to the
reference and, if picked up in the signal, will add or
subtract from the actual signal and cause errors in
the measurement. Typical sources of synchronous
noise are ground loops between the experiment,
detector and lock-in, and electronic pick up from
the reference oscillator or experimental apparatus.

Many of these noise sources can be minimized
with good laboratory practice and experiment
design. There are several ways in which noise
sources are coupled into the signal path.

Capacitive coupling

An AC voltage from a nearby piece of apparatus
can couple to a detector via a stray capacitance.
Although C

stray

may be very small, the coupled

noise may still be larger than a weak experimental
signal. This is especially damaging if the coupled
noise is synchronous (at the reference frequency).

We can estimate the noise current caused by a
stray capacitance by,

where

ω

is 2

π

times the noise frequency, V

noise

is

the noise amplitude, and C

stray

is the stray

capacitance.

For example, if the noise source is a power circuit,
then f = 60 Hz and V

noise

= 120 V. C

stray

can be

estimated using a parallel plate equivalent capaci-
tor. If the capacitance is roughly an area of 1 cm

2

at a separated by 10 cm, then C

stray

is 0.009 pF.

The resulting noise current will be 400 pA (at
60 Hz). This small noise current can be thousands
of times larger than the signal current. If the noise
source is at a higher frequency, the coupled noise
will be even greater.

If the noise source is at the reference frequency,
then the problem is much worse. The lock-in
rejects noise at other frequencies, but pick-up at
the reference frequency appears as signal!

Cures for capacitive noise coupling include:

1) Removing or turning off the noise source.

2) Keeping the noise source far from the

experiment (reducing C

stray

). Do not bring

the signal cables close to the noise
source.

3) Designing the experiment to measure volt-

ages with low impedance (noise current
generates very little voltage).

4) Installing capacitive shielding by placing

both the experiment and detector in a
metal box.

Inductive coupling

An AC current in a nearby piece of apparatus can
couple to the experiment via a magnetic field. A
changing current in a nearby circuit gives rise to a
changing magnetic field which induces an emf
(dØ

B

/dt) in the loop connecting the detector to the

experiment. This is like a transformer with the
experiment-detector loop as the secondary
winding.

EXTERNAL NOISE SOURCES

i

=

C

stray

dV

dt

= ω

C

stray

V

noise

Detector

Stray Capacitance

Noise

Source

Experiment

Detector

Noise

Source

Experiment

B(t)

Advertising