Chapter 3, Operation – Lakeshore Learning Materials 647 User Manual

Page 27

Advertising
background image

Lake Shore Model 647 Magnet Power Supply User’s Manual

Operation

3-1

CHAPTER 3

OPERATION

3.0 GENERAL
This chapter covers seven areas: MPS Front Panel (Paragraph 3.1), Power Up (Paragraph 3.2), Setting

Current on an MPS with Manual PSH Control (Paragraph 3.3), Setting Current with Automatic PSH Control

(Paragraph 3.4), Instrument Setup Screens (Paragraph 3.5), Function Menus (Paragraph 3.6), and an

Automatic Persistence Control Example (Paragraph 3.7).

Setup the MPS software with the MPS output terminals shorted together as in section 3.2. This ensures that

while the user learns MPS operation, an inadvertent keystroke causes no damage to the magnet.

3.1 THE MPS FRONT PANEL
Figure 3.1 below shows the MPS Front Panel. The up or down Display Cursor Control keys move the line

indicator to the line containing a value to be changed. Use either the Data Entry Keypad to enter the desired

value or the up or down Numeric Entry keys to increment or decrement the value. The Enter key accepts the

update, while the Esc key discards the change and returns to the prior value. Move the cursor to a particular

digit with the right and left Display Cursor Control keys and change it with the Numeric Entry keys.

Function

Menu

Next

Menu

Normal
Display

Cursor

Mode

Local

Remote

Output

Inhibit

7

8

9

Esc

4

5

6

1

2

3

0

+/–

Enter

Data Entry

Functions

Display Menu Select Keys

Display Function Keys

Display Cursor Control Key

Data Entry Keypad

Interface Keys

Enter Key

accepts a change

Esc Key returns to prior value

Numeric Entry

Increment and Decrement Keys

Output

Inhibit Key

Figure 3-1

Front Panel Numeric Entry Keys

3.2 POWER

UP

Read and follow instructions in Paragraphs 2.1 - 2.4.3 and the Forward safety recommendations before

applying power to the MPS. Do not connect the MPS to the magnet yet. Short the output terminals together

with a #4 gauge or larger cable to protect the magnet against incorrect configurations.

Turn on the MPS. It requires approximately 2 seconds for initialization. Initially, all front panel annunciators

come on and the alarm sounds for a short time. Within 1 second, the Fault and Persistent Switch Heater On

annunciators and the alarm turn off. If the MPS detects a high or low AC line fault, it blinks the front panel

Fault annunciator and turns off the input circuit breaker. If this occurs, verify that the AC source matches the

line voltage listed on the MPS rear panel.

Initially, the entire display clears and the alarm sounds for a short time. The MPS initializes itself and displays

the model identification. The Normal Display screen appears with a blinking asterisk indicating each update

when the unit is in normal operation.

CAUTION: Set magnet parameters according to manufacturer's specifications. Failure to do so may

damage the magnet and threaten user safety.