Lakeshore Learning Materials 647 User Manual

Page 55

Advertising
background image

Lake Shore Model 647 Magnet Power Supply User’s Manual

Options & Accessories

5-5

20. Convert DVM voltage reading to current. A

reading of +0.512345V equals +51.2345A. Send

actual current with ‘CALPL+xxx.xxxx’. If actual

current is +51.2345A, send ‘CALPL+51.2345’.

The MPS determines the positive current

calibration constant.

21. After the MPS determines the positive current

calibration constant, it automatically sets output

current to the same value with the opposite sign.

Verify that the DVM reads -0.05V ±10%. Allow

MPS output to settle for 2 minutes.

22. Convert voltage read from the DVM to current. A

reading of -0.0487655V equals -48.7655A. Send
actual current with the command ‘CALMN-
xxx.xxxx’. If actual current is 48.7655A, send

‘CALMN-48.7655’. The MPS determines the
negative current calibration constant.

23. The MPS stores the zero, positive, and negative

current calibration constants then sets output
current to 0. Verify a DVM reading of 0V
±0.00005V before continuing the calibration.

24. Use IMAX and STPS commands to restore

IMAX and output current step to Step 7 values.

25. Turn off MPS and disconnect calibration loads.

26. Restore original Control Bus Serial Interface

settings.

27. Disable MPS calibration. Turn OFF (down) CAL

AND ID switch 8.

5.4 PERFORMANCE

TEST

Performance tests verify proper MPS operation into

a resistive load without testing all specified

parameters. The tests include no troubleshooting

information, but can provide information that may

localize faults. If any tests fail, contact authorized

service personnel.

WARNING: For service, return the MPS to Lake

Shore or a factory representative. Lake Shore

Cryotronics, Inc. cannot be held liable for injury

or death of personnel attempting unauthorized

MPS repairs.

Test Equipment Required

1. Current monitor shunt, 100 mV output, rated at

150 A . Manufacturers typically stock this shunt

value. If a different value is used, adjust

readings given in the procedure.

2. Digital Volt Meter (DVM) capable of reading DC

voltage from 100 mV with 0.1 mV accuracy to 35

V with 0.1 V resolution.

Performance Test Procedure

The test procedure includes both front panel and

remote operation. Execute the performance tests in

the order given. Note any non-compliance.

Test Setup

1. Connect input power to the MPS as outlined in

Sections 2.5 through 2.5.3 of this manual.

2. Disconnect any leads from MPS output

terminals.

3. Set Programming Mode Switches on the MPS

rear panel(s) to the INT (up) position. Set the

CAL and ID switch 7 OFF (down).

4. Turn ON MPS. Initially, the display clears and

the alarm sounds for a short time. Within two

seconds, the Normal Display appears.

Output Voltage Verification

5. From the MPS front panel, select the Instrument

Setup (INSTR. SETUP) menu and change com-

pliance voltage to 5 V. Press the Normal Display

key to return to the Normal Display screen.

6. Set output current to +1 A. Use the DVM to read

the voltage across the MPS unit(s) output

terminals. The reading should be +5 V ±0.3 V.

7. Set output current to 1 A. The DVM should read

5 V ±0.3 V.

8. Turn OFF MPS.

Output Current Verification

9. Use AWG #4 wire to connect the current monitor

shunt between the MPS +OUT and -OUT

terminals. Connect DVM across the shunt

monitoring terminals. Note DVM offset reading.

10. Turn ON MPS. Within two seconds, the Normal

Display appears.

11. Select the Instrument Setup (INSTR. SETUP)

menu and change compliance voltage to 2

Volts. Press the Normal Display key to return to

the Normal Display screen.

12. Wait 1 minute for MPS output to settle. The

DVM should read the same as recorded for Step

9 ±0.01 mV.

13. Set output current to +50 Amps The DVM

should read about +33 mV. After about 1

minute, the DVM should read +33.333 mV

±0.117 mV (not including offset from Step 9).

Advertising