Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Error Messages & Troubleshooting

5-4

5.2

Factory Default Settings

Below is a list of factory-presets. To reinitialize the MPS to factory presets at any time, hold the Esc key for

about 15 seconds.

NOTE: If the MPS is part of a superconducting magnet system, the system supplier may change some of the

parameters below to optimize MPS operation to the superconducting magnet it works with.

Automatic Mode Screen – This screen is unavailable if PSH option is not installed

MGT START: 0.000A

– Magnet Start

MGT END: 0.000A

– Magnet End

MGT RATE: 1.000A/S

– Magnet Rate

PERS RATE: 1.000A/S

– Persistence Rate

LAST END: 0.000A

– Ending current from last magnet charge

PERSISTENCE: <ON>

– Persistence Mode Status

RUN: ¥

– Ramp Start Toggle

Automatic Mode Delays Screen

RUN UP: 60 S

– Run Up

PSH OFF: 60 S

– PSH Off

RAMP END: 60 S

– Ramp End

PSH ON: 60 S

– PSH On

Manual Mode Screen

CURRENT: +0.000A

– Current Setting

COMPLIANCE: +1.000V

– Compliance Voltage Setting (Cannot be changed in this screen)

RAMP RATE: +1.000A/S

– Ramp Rate

RAMP ENABLE: <OFF>

– Ramp Enable Off

PERSISTENCE: <ON>

– Persistence Status (Not present if PSH option not installed)

Instrument Setup Screen

FLD ENABLE: <OFF>

– Magnetic Field Enable Off

B UNITS: kG

– Magnetic B-Field Units

kG/A: ±0.100

– Field per Amperes

IMAX SET: ±50.000A (620), ±125.000A (622), ±155.000A (623), ±72.000A (647) – Maximum Current

COMPLIANCE: ±5.000V (620), ±30.000V (622/623), ±32.000V (647),

– Maximum Compliance Voltage

(Retains Last Setting On Power Down)

Step Limit/Zero Screen

CURRENT STEP LIMIT:

STATUS: <ON>

– Output Current Step Limit Status

I STEP: +10.000A

– Output Current Step Limit

ZERO:

STATUS: <OFF>

– Output Current Zero Status

I ZERO: +0.000A

– Output Current Zero Value

I MODE: INTERNAL

– Output Current Programming Mode (Display Only)

Interface Setup Screen (If Present)

IEEE ADD: 12

– IEEE 488 Address

TRMS/EOI: (CR LF) EOI – Terminators (Carriage Return & Line Feed) and EOI On

RS-232C: 1200,N,8,1

– Serial I/O Definition (9600 Baud, No Parity, 8 data bits, 1 stop bit)

CB DEF: 9600,o,7,1

– Control Bus Definition (9600 Baud, Odd Parity, 7 data bits, 1 stop bit)

CB ADD: 0

– Control Bus Address

VW ANGLE: 5

– Viewing Angle (5 = Straight On)

MF ID: 622 (1)

– Mainframe ID Number

Liquid Helium Level Screens (If Present)

READING: 00.00 <OFF>

– Level Reading and Status

UNITS: in (E)

– Level Probe Units – Inches (English)

NEXT RDG: 00:00

– Next Reading (HH:MM)

PERIOD: 01.00

– Period (HH:MM)

ALARM: +0.0<OFF>

– Level Alarm and Status

ZERO REF: +0.0<OFF>

– Level Zero Reference and Status

Field Screen (If Present)

Persistent Switch Heater (PSH) Screen (If Present)

FIELD PROBE SETTING:

HEATER: 50mA

– PSH Current

UNITS: kG

– Field Units in Kilogauss

STATUS: <OFF>

– PSH Status

CALIBRATE PROBE:

LAST END: 0.000A – Ending current from last magnet charge

mV/kG: +0.800

– Field Probe Sensitivity