Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

5-8

Error Messages & Troubleshooting

5.5 REAR PANEL CONNECTOR DETAIL

Figure 5-1 MPS Rear Panel Connectors

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18

+

-

+

-

+

-

OVP Im Vm m Vp +Vs -ls

lp +ls -S +S

CAL AND ID

V

l

MODE

INT

V

l

EXT

RI

FLT

ON

10K

MPS Rear Panel Connectors

10K

8 7 6 5 4 3 2 1

Active

Open

Table 5-1B Rear Panel CAL

AND ID Switch Definitions

Switch Function

1 - 3

LSB Multi-

Mainframe Address

4 Multi-Mainframe
5 Dither
6 Not

Used

7 Inhibit

Delay

8 Calibration

Table 5-1A Rear Panel Connector Definitions

TERMINAL LABEL

DEFINITION

1
2

RI+
RI–

Remote Inhibit – Active low, TTL-compatible
input to remotely disable the output (force the
output settings to 0 A and 1 V).

3
4

FLT +

FLT–

Fault Indicator – Normally open contact closure
to indicate a fault condition has occurred.
Contact rating: 0.25 A resistive at 100 VDC, 3 W,
25 VA.

5
6

ON+
ON–

ON Indicator – Contact closure to indicate when
the front panel circuit breaker is in the ON
position. Contact rating: 0.25 A resistive at 100
VDC, 3 W, 25 VA.

7 Reserved

None


8


OVP

Overvoltage Protection Enable – Analog signal
connected in parallel to other MPS Units to
remotely activate the output overvoltage
protection circuit.

9 Im

Output Current Monitor – Voltage output from Im
to GND(M) is ±0.01 V/A.

10 Vm

Output voltage monitor – Voltage output from Vm
to GND(M) is ±0.01 V/V.

11 m

Monitor and program ground. GND(M).

12 Vp

Factory Use Only

13 +Vs

Factory Use Only

14 –Is

Enable external output current programming via
MODE switch.

15 Ip

Voltage input from Ip to GND(M) is ±0.01 V/A.


16


+Is

A 10 k

potentiometer from +Is to –Is with center

tap to Ip produces the minimum voltage for full
scale current output (±1.25 V for Model 622 or
±1.55 V for Model 623). Voltage applied to Ip
sums with internal current programming voltage.

17
18

–S
+S

Remote voltage sense correction. Correction for
load lead drops of up to 0.5 V per lead.

Figure 5-2 Serial

Interface Connector

1 2 3 4 5 6

Table 5-2 Serial Interface

Connector Definition

PIN

RS-232C

FUNCTION

RS-485

FUNCTION

1

No

Function

No

Connection

2

Receive

Data

A

3

Signal

Ground

No

Connection

4

Signal

Ground

No

Connection

5

Transmit

Data

B

6

No

Connection

No

Connection

Advertising
This manual is related to the following products:

623, 620