Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Error Messages & Troubleshooting

5-9

5.6 IEEE-488 INTERFACE CONNECTOR (WHEN MODEL 6224 OPTION IS PRESENT)
Connect to rear MPS IEEE-488 Interface connector with cables specified in the IEEE-488-1978 standard

document. The cable has 24 conductors with an outer shield. The connectors at each end are 24-way

Amphenol 57 Series (or equivalent) with piggyback receptacles to daisy-chain multiple devices. Secure

connectors in the receptacles by a pair of captive locking screws with metric threads.

The total length of cable allowed in a system is 2 meters for each device on the bus, or 20 meters total for a
maximum of 15 devices. The figure below and the table to the right shows the IEEE-488 Interface connector
pin location and signal names.

12

11

10

9

8

7

6

5

4

3

2

1

24

23

22

21

20

19

18

17

16

15

14

13

SH1 AH1 T5 L4 SR1 RL1 PP0 DC1 DT0 C0 E1

PIN SYMBOL

DESCRIPTION

1
2
3
4
5
6
7
8
9

10
11
12
13
14
15
16
17
18
19
20
21
22
23
24

DIO1
DIO2
DIO3
DIO4
EOI
DAV
NRFD
NDAC
IFC
SRQ
ATN
SHIELD
DIO5
DIO6
DIO7
DIO8
REN
GND 6
GND 7
GND 8
GND 9
GND 10
GND11
GND

Data Input/Output Line 1
Data Input/Output Line 2
Data Input/Output Line 3
Data Input/Output Line 4
End Or Identify
Data Valid
Not Ready For Data
Not Data Accepted
Interface Clear
Service Request
Attention
Cable Shield
Data Input/Output Line 5
Data Input/Output Line 6
Data Input/Output Line 7
Data Input/Output Line 8
Remote Enable
Ground Wire – Twisted pair with DAV
Ground Wire – Twisted pair with NRFD
Ground Wire – Twisted pair with NDAC
Ground Wire – Twisted pair with IFC
Ground Wire – Twisted pair with SRQ
Ground Wire – Twisted pair with ATN
Logic Ground

Figure 5-3. IEEE-488 Interface Connector