Lake Shore Model 620/622/623/647 Magnet Power Supply User’s Manual

Options & Accessories

6-7

IBS?

Magnetic Field Summary Query

Input:

IBS?

Returned: [IOUT], [BFLD], [STB]
Remarks:
Returns 23 characters plus up to two terminators.

LALM

Program Level Alarm

Input:

LALM [alarm] where [alarm] is the level value between 0 and ±999.9 that activates the
alarm. The value is truncated (not rounded) to the tenths place.

Remarks: The initial condition is 0. The alarm value uses the existing units. Conversion is made if units

are changed.

LALM?

Level Alarm Query

Input:

LALM?

Returned: The value returned is a number between ±999.9000.
Remarks: Nine characters plus up to two terminators are returned.

LALMS

Program Level Alarm Status

Input:

LALMS [status] where [status] = 0 for OFF, or 1 for ON.

Remarks: The initial condition is 0.

LALMS?

Level Alarm Status Query

Input:

LALMS?

Returned: 0 or 1. 0 indicates that the level alarm has not been reached. 1 indicates that the level is

below the alarm value and the alarm has been activated.

Remarks: Returns one character plus up to two terminators.

LEVL

Read LHe Level

Input: LEVL

[status] where [status] = 1 to read the level, or 0 not to read it.

Remarks: Reading the level requires a minimum of 5 seconds to update the level reading.

LEVL?

LHe Level Query

Input: LEVL?
Returned:
A number between ±999.9000.
Remarks: Returns nine digits plus up to two terminators.

LUNI

Program LHe Level Units

Input: LUNI

[units]

where [units] = "%" for percentage, "E" for English units (inches), or "M" for

metric units (centimeters).

Remarks: The initial condition is E (inches). Make sure there is a space between the command and the

units parameter.