2 cleaning the cabinet filters, 3 cabinet – Pioneer Q150 User Manual

Page 15

Advertising
background image

14

4. The oxygen concentration should be checked every 90 days for units without optional oxygen

sensors. The oxygen concentration should be checked every 6 months for units with optional
oxygen sensors.

10.2 CLEANING THE CABINET FILTERS

NOTE: All concentrator models have one cabinet filter located at rear side of the cabinet.

(FIG. 6)

FIGURE 6

CABINET FILTER LOCATION

1. Remove the cabinet filter and

CLEAN AT LEAST ONCE A WEEK depending on environmental

conditions.

(FIG.7)

2. Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.

(FIG. 7)

3. Dry the filters thoroughly before reinstallation.

FIGURE 7

REMOVE THE CABINET FILTERS


10.3 CABINET

Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.

Advertising