V I I . C A R E A N D M A I N T E N A N C E

All wheel bearings are prelubricated and sealed. They require no subsequent lubrication.

The body shroud has been sprayed with a clear sealant coating. You can apply a light coat of car wax to help

it retain its high-gloss appearance.

Check all electrical connections. Make sure they are tight and are not corroded. Batteries must sit flat within

the battery tray, with the battery terminals facing inward, toward each other. Refer to the battery wiring label

for the correct wiring layout.

WARNING! Even though the power chair has passed the necessary testing

requirements for ingress of liquids, you should keep electrical connections away from

sources of dampness, including direct exposure to water or bodily fluids and

incontinence. Check electrical components frequently for signs of corrosion and

replace as necessary.

Daily Checks

With the controller turned off, check the joystick. Make sure it is not bent or damaged and that it returns to the

neutral position when you release it. Check the rubber boot around the base of the joystick for damage. Visually

inspect the boot. Do not handle or try to repair it. See your Quantum Rehab Provider if there is a problem.

Visually inspect the controller harness. Make sure that it is not frayed, cut, or has any wires exposed. See your

Quantum Rehab Provider if there is a problem.

Check for flat spots on solid tires. Flat spots could adversely affect stability.

Inspect the armrests for loose hardware, stress points, or damage. See your Quantum Rehab Provider if there

is a problem.

Weekly Checks

Disconnect and inspect the controller from the power base. Look for corrosion. Contact your Quantum Rehab

Provider if necessary.

Ensure that all parts of the controller system are securely fastened to your power chair. Do not overtighten any screws.

Check for proper tire inflation. Maintain but do not exceed the psi/bar/kPa air pressure rating indicated on each

tire. If a tire does not hold air, see a Quantum Rehab Provider for replacement of the tube.

Check the brakes. This test should be carried out on a level surface with at least 3 feet (1 meter) of clearance

around your power chair.

To check the brakes:

1. Turn on the controller and turn down the speed level of your power chair.

2. After one second, check the battery condition meter. Make sure that it remains on.

3. Slowly push the joystick forward until you hear the electric brakes click. Immediately release the joystick. You

must be able to hear each electrical brake operating within a few seconds of joystick movement. Repeat this

test three times, pushing the joystick backwards, then left, and then right.

Monthly Checks

Check for drive tire wear. See your Quantum Rehab Provider for repair.

Check the caster wheels for wear. Replace them as necessary.

Check the caster forks for damage or fluttering which indicates that they may need to be adjusted or have the

bearings replaced. See your Quantum Rehab Provider for repair.

Keep your power chair clean and free of foreign material, such as mud, dirt, hair, food, drink, etc.

Yearly Checks

Take your power chair to your Quantum Rehab Provider for yearly maintenance, especially if you use your power

chair on a daily basis. This helps ensure that your power chair is functioning properly and helps prevent future

complications.