Device Errors 

Error or symbol 

Cause 

Resolution 

Device shows  
"E 
  E" vertically in two lines 

Cuff is over‐inflated.  

Remove arm cuff.  Refer to “Taking a 
Measurement” in instruction manual. 
 

Device shows “E” 

Monitor could not detect 
pulse wave. 
 
Air plug is not connected 

Take another measurement and remain 
still until the measurement is complete. 
 
Check air plug connection.   

Device shows “Er” 

A malfunction has occurred.   Contact Omron Customer Support at 

www.omronhealthcare.com

 

Device shows “Battery 
picture with a cross mark” 

Batteries are worn 

Replace the four batteries. 
Refer to “Battery Installation 

No Power 
 
No Display appears on the 
unit 
 

Power off 

Check that the device is turned on.  
 
Check the battery installation for proper 
placement of the battery polarities. 
 
Replace all four batteries with new ones. 
 