Omega Speaker Systems M1760 User Manual

Bcm-1

Advertising
background image

DESCRIPTION

The OMEGA

®

BCM-1 Bridge Completion Module provides

a convenient means of completing the Wheatstone Bridge

circuit used for strain gage measurements. The module

can be used for quarter bridge measurements with 120 or

350 ohm gages or for half bridges with gages of any

resistance. Quarter bridges with two or three wire hookup

can be accommodated. In addition, a bridge balance

adjustment potentiometer is provided to zero the bridge

output. Screw terminals are provided for strain gage,

excitation, and output signal connections.

The BCM-1 is constructed using precision resistors with 5

PPM/°C temperature coefficient for maximum temperature

stability. The unit is completely encapsulated for use in

rugged environments and can be mounted by using the

screw holes provided in the baseplate.

MOUNTING

The BCM-1 module is equipped with a baseplate which

provides holes for screw mounting. Where required, it can

also be glued or secured in place with double-sided tape

or Velcro.

SPECIFICATIONS

Maximum Excitation

120 Ohm Bridge:

10 Vdc

350 Ohm Bridge:

16 Vdc

Temperature Limits:

-20 to 80°C

(-4 to 176°F)

Temperature Effects:

±1.5 microvolts/volt/°C

Zero Adjust:

±6 mV/V

Resistor Tolerance:

±0.1%

Resistor TEMPCO:

5 PPM/°C

Dimensions:

3.0" W x 1.25" H x 1.0" D

(50.8 mm x 31.8 mm x

25.4 mm)

Mounting Holes:

0.175" holes (2) on

2.5" center

BCM-1 HOOKUP

Connections to the BCM-1 are made using the screw

terminals on the top of the unit using stripped wires. One

side has four terminals used for excitation and signal

output connections. The excitation voltage is connected to

the outer terminals labeled +EX and -EX. The readout

device is connected to the two center terminals which are

labeled + and - Vout.

Connection of the strain gages to the BCM-1 depends on

the particular application. For the proper hookup, refer to

Figures 1-3.

+BRG

+BRG

+SIG

120

350

-BRG

* FOR 2 WIRE

GAGES INSTALL

JUMPER FROM

+SIG TO

120 TERMINAL

FIG.1 1/4 Bridge 120 Hook-Up

BCM-1

+BRG

+BRG

+SIG

120

350

-BRG

* FOR 2 WIRE

GAGES INSTALL

JUMPER FROM

+SIG TO

350 TERMINAL

FIG.2 1/4 Bridge 350 Hook-Up

BCM-1

+BRG

+BRG

+SIG

120

350

-BRG

FIG.3 1/2 Bridge Hook-Up

BCM-1

*

*

Figure 1. 1/4 Bridge 120

Ω Hook-up

Figure 2. 1/4 Bridge 350

Ω Hook-up

Figure 3. 1/2 Bridge Hook-up

BCM-1

Bridge Completion Module

INSTRUCTION

SHEET

M1760/1106

Shop online at: omega.com e-mail: [email protected]

For latest product manuals: omegamanual.info

Advertising
This manual is related to the following products: