Humidifier performance – ResMed S9 H5i User Manual

Page 10

Advertising
background image

English

6

Notes:

The manufacturer reserves the right to change these specifications without notice.

The temperature and relative humidity settings displayed for ClimateLine or ClimateLine

MAX

are not

measured values.

Check with your clinician/service provider before using the SlimLine air tubing with devices other
than the S9 or H5i.

Humidifier performance

The following settings have been tested at 22°C ambient temperature:

Cleanable water tub

Injection molded plastic, stainless steel and silicone seal

Standard water tub

Injection molded plastic, aluminium and thermoplastic
elastomer

Operating temperature

+5ºC to +35ºC

Operating humidity

10–95% non-condensing

Storage and transport temperature

-20ºC to +60ºC

Storage and transport humidity

10–95% non-condensing

ClimateLine air tubing

Flexible plastic and electrical components, 2 m, 15 mm
inner diameter

ClimateLine

MAX

air tubing

Flexible plastic and electrical components,1.9 m, 19 mm
inner diameter

Heated tubing temperature cut-out

 41°C

IEC 60601-1 classification

Class II (double insulation), Type BF

Electromagnetic compatibility

Please refer to the S9 Information Guide for details

CPAP mask

pressure, cm H

2

O

RH output %

Nominal system output AH

a

, BTPS

b

a.

AH - Absolute Humidity in mg/L.

b. BTPS - Body Temperature Pressure Saturated.

Setting 3

Setting 6

Setting 3

Setting 6

4

90

100

10

18

10

95

100

11.5

21

20

95

100

11

18

25

100

100

12

13.5