Calibrating the compass, 3 calibrating the compass, 1 setting the local declination – SUUNTO Advizor User Manual

Page 50

Advertising
background image

44

6.2.1 Setting the Local Declination

1. PRESS the [Select] button and hold for 2 seconds. Located in Field 1, “OFF” will begin to flash.

2. PRESS either the [+] or the [-] button to change the declination direction in Field 1.

3. At the direction desired, PRESS the [Select] button to move to the next setting. Located in Field 2, the degrees will

begin to flash.

4. PRESS the [+] button to scroll the degrees upward or PRESS the [-] button to scroll the degrees downward.

5. At the degrees desired, PRESS the [Mode] button to accept the changes and exit the setup program.

Note: If the user does not press any button for 1 minute in the setup mode, the display will automatically exit setup.

Local Declination is now set.

6.3 CALIBRATING THE COMPASS

A compass calibration should be done whenever the Wristop Computer has been subject to close proximity of magnetic
sources, extreme cold weather, every time the battery is replaced or it is suspected that other environmental conditions
have effected the compass readings.

Note: We also recommend calibrating the compass before using the compass the first time, as well as when heading
outdoors to use the compass.

To begin the calibration:

1. PRESS the [Select] button twice when in the main compass mode. Field 1 displays the text “CMP”. Field 3 displays

the text “CAL” (Fig. 56).

2. PRESS the [Select] button and hold for 2 seconds. Located in Field 2, the text “PUSH” will begin to flash (Fig. 57).

3. PRESS the [-] button to start calibration .

4. In Field 2, 360° appears and the Outer Circumference displays all segments lit (Fig. 58). When this occurs, hold the

Wristop Computer level and start to slowly turn the device around a full circle in a level position. The direction of turning
does not matter. The lit segments will be turned off as the turning proceeds. When the first full circle is completed,
the instrument will inform the user within one minute of whether the calibration was successfull, displaying the text
“dOnE” in Field 2 (Fig. 59), or if the process needs to be repeated, displaying “PUSH” and “-” in Field 2.

Advertising
This manual is related to the following products: