105

104

IF...

the following screen is displayed.

REASON

The temperature is outside the operating range of
the test strips (57°-104°F or 14°-40°C).

ACTION

The result you have obtained may not be
accurate.
A temperature symbol will appear below
your result. Move to an area between 57° and
104°F, wait 5 minutes and repeat the test. Do not
artificially heat or cool the meter.

▲ Example result

Blue circle placed
over screen captures
prints PMS 287.

IF...

the following screen is displayed.

REASON

The meter is turned on and the memory is full with
100 values. MEM flashes.

ACTION

If a new result is added, the oldest result is deleted.