109

108

IF...

the following screen is displayed.

REASON

The meter has failed its internal system check.

ACTION

Remove the Chek

Strip. Turn the meter off and

repeat the glucose control check. See

Glucose

Control Testing

(page 27).

If this error display still

appears, testing cannot take place. Call Accu-
Chek

SM

Customer Care at 1-800-858-8072.

IF...

the following screen is displayed.

REASON

The test strip may be damaged or the test was not
performed correctly.

ACTION

If this display appears

before blood is placed on the

strip, remove the test strip and reinsert. If this
display appears

after blood is applied to the strip,

there is an incorrect amount of blood on the strip.
Repeat the test with a new test strip.

Troubleshooting Guide