Activeforever Mettler Sonicator Plus 930 2-Channel Combination Therapy Device User Manual

Sonicator, Plus 930 specifications, General specifications

Advertising
background image

Sonicator

®

Plus 930 Specifications

General Specifications:

Input:

90–240 VAC, 50–60 Hz, 2.3 Amp Nom.

Certification:

The Sonicator Plus 930 complies with the ultrasound
performance standards set forth in the Code of Federal
Regulations, Title 21 (Food and Drugs), Part 1050.10

ETL and C-ETL Listed:

Model ME 930 (9801427)

Classification:

Protective Class I Equipment

Type BF Equipment

Enclosed equipment without protection against ingress of
water.

Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air or with nitrogen
oxide

Year 2000 Compliant

Yes

U.S. Patent Numbers:

4,966,131 and 5,095,890

Weight: 8.7

pounds

4.0 kg

Dimensions:

6 in (H) x 12 in (W) x 12 in (D)
15.2 cm (H) x 30.5 cm (W) x 30.5 cm (D)

Operating Temperature:

+50°F to +104°F
+10°C to +40°C

Humidity:

Operating, 30% to 75% Relative Humidity at 104°F
Nonoperating, 5 to 95% Relative Humidity, non-
condensing

Storage Temperature:

-40°F to 167°F
-40°C to 75°C

Timer Accuracy:

±0.5 minutes for times less than 5 minutes
±10% for times from 5 to 10 minutes
±1.0 minute for times greater that 10 minutes

Maximum Treatment Time:

60 minutes–electrical stimulation
30 minutes–ultrasound or combination therapy

Treatment Timer:

Treatment time counts down to zero when a time is set, or
up to 60 or 30 minutes when no time is set. The digital
timer indicates time in minutes and seconds. The timer
also indicates the remaining or elapsed treatment time
during the “Hold” period.