CALIBRATION

CHAPTER 5


4.

Enter the following code into the display using the Cursor Keys:

5111. Then press

( Left (

←), Right (→) to select digit ; Up (↑), Down (↓) = increment/decrement value ;

Enter (

↵) to accept )

Calibrate tester

Using a laboratory millivolt meter, measure the voltage across pin 2 (ground) and 19 (Gage

Excitation) of the calibration fixture.


The

display

reads:


Example:

3.0211 volts


Program the measured voltage into the SURETEST Base Unit using the Cursor Keys. (Use Up (

↑),

Down (

↓), Left (←) and Right (→) Cursor Keys to match the SURETEST display to the millivolt-

meter). Then Press ENTER (

↵) to accept.

5.

Next, the display reads:

Short pins 5 and 6 together on the calibration fixture. Then hit Enter (

↵) key.


6.

Next, the display reads:


Remove the short between pins 5 and 6 of the calibration fixture.

Using a laboratory millivolt meter, measure the voltage across pins 5 and 6.