A - Specifications

144

Supplemental Characteristics

Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical
performance determined either by design or type testing.

Table A-2. Supplemental Characteristics

Parameter

Agilent 66111A

Agilent

66311B/D,

Agilent 66309B/D

output 1 only

Agilent

66309B/D

output 2 only

Output Programming
Range

Voltage:

Current:

OVP:

0 – 15.535 V
0 – 3.0712 A

22 V

0 – 12.25 V

0 – 1.52 A

NA

Average Programming
Resolution

Voltage:

Current:

OVP:

3.75 mV
0.75 mA

100 mV

14 mV

1.75 mA

NA

OVP Accuracy

2.4 % + 240 mV

NA

Average Current
Measurement Resolution

High Range:

Low Range:

3.5 mA

NA

213

µ

A

0.6

µ

A

55

µ

A

NA

Maximum Current
Measurement

7 A

1.8A

Sink Current

- 2 A @ 7.5 V

1

- 0.03 A @ 7.5 V

Programming Accuracy
Temperature Coefficient
(change/C

°

)

Voltage:

Current:

OVP:

0.01% + 0.5 mV

0.01% + 15

µ

A

0.015% + 4 mV

0.03% + 1.5 mV
0.03% + 0.4 mA

NA

Readback Accuracy
Temperature Coefficient
(change/C

°

)

Voltage:

Current (dc):

Current (acdc):

Current (low range):

0.01% + 300

µ

V

0.02% + 15

µ

A

NA
NA

0.01% + 300

µ

V

0.02% + 15

µ

A

0.05% + 120

µ

A

0.01% + 0.3

µ

A

0.02% + 1.5 mV
0.02% + 0.4 mA

NA
NA

Drift

2

Voltage:

Current: +

0.01% + 1 mV

0.01% + 30

µ

A

0.03% + 3 mV

0.03% + 0.8 mA

Remote Sense Capability
(voltage per load lead)

up to 4 V

3

up to 4 V

4

Output Voltage Rise Time
Output Voltage Fall Time

10% to 90%:
90% to 10%:

< 200

µ

s

< 200

µ

s

< 200

µ

s

< 35 ms

Output Voltage Settling
Time

5

0 to rated voltage:
rated voltage to 0:

0.5 ms

1 ms

NA
NA

Dynamic Measurement
System

6

Buffer Length:

Sample Rate Range:

NA

0

4096 points

15.6

µ

s

31200s

NA

1

The sink current decreases linearly from 2.8A @ 0 V to 1.2 A @ 15 V. Sink current does not track the programmed current.

2

Following a 30 minute warm-up, the change in output over 8 hours, under constant ambient, load and line conditions.

3

Add 2 mV to the load regulation for each 1 V change in the positive output lead.

4

Add 3 mV to the load regulation for each 1 V change in the negative output lead.

5

To settle within 1 LSB or 3.75mV of the final value for Output 1.

6

For a pulse waveform, the accuracy of any individual data point in the buffer is dependent on the rise time of the pulse. For a

current pulse of 1.4A with a rise time constant of 50

µ

s, the error in measurement of a single data point during the pulse rise time

is typically 10mA.