10 kV Voltage Power Supply Characteristics

28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY

http://www.keithley.com

Specifications are subject to change without notice.
SPEC-2290-10 Rev. A / December 2013

Page 1 of 3

CONDITIONS

This document contains specifications and supplemental information for the Model 2290-10 High Voltage
Power Supply. Specifications are the standards against which the Model 2290-10 is tested. Upon leaving
the factory, the Model 2290-10 meets these specifications. Characteristics, supplemental characteristics,
and typical values are not warranted, apply at 23 °C ± 5 °C, < 70% relative humidity, and are provided
solely as useful information.

CHARACTERISTICS

Voltage range:

Output voltage

1

Maximum output current

+100 to +10,000 V DC

1.000 mA DC

Voltage set accuracy: ±0.06% of full scale

Voltage display accuracy: voltage set accuracy ±1 V, typical (±2 V, maximum)

Voltage resolution: 1 V (set and display)

Voltage limit range: 0 to 100% full scale

Voltage regulation

2

:

Line: 0.001% for ±10% line voltage change

Load: 0.04% for 100% load change, typical

Output ripple (300Hz – 300 kHz):

0.01% of full scale, V RMS

Rise time (from high voltage turn on to final value under full load)

3,

4

:

<6 seconds to within 1 V of the final value

Discharge time (to 0 V under fulll load)

3, 4

:

<1 second to under 10 V

Discharge time (no load)

3

:

<6 seconds (to <1% of full scale voltage with no load, typical)

Output Stored Charge: <20 µC maximum

1

The output voltage can be programmed to voltages below 100 V, but accuracy below 100 V is not specified.

2

Regulation specifications apply for greater than 100 V

3

Current limit set to 105% of full scale.

4

Under resistive load.