Atec Keithley-2290-5 User Manual

5 kv voltage power supply characteristics

Advertising
background image

Model 2290-5

Keithley Instruments, Inc.

5 kV Voltage Power Supply Characteristics

28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY

http://www.keithley.com

Specifications are subject to change without notice.
SPEC-2290-5 Rev. A / December 2013

Page 1 of 3

CONDITIONS

This document contains specifications and supplemental information for the Model 2290-5 High Voltage
Power Supply. Specifications are the standards against which the Model 2290-5 is tested. Upon leaving the
factory, the Model 2290-5 meets these specifications. Characteristics, supplemental characteristics, and
typical values are not warranted, apply at 23 °C ± 5 °C, < 70% relative humidity, and are provided solely as
useful information.

CHARACTERISTICS

Voltage range:

Output voltage

1

Maximum output current

Conditions

50 to +5000 V dc

5.000 mA DC

No filter

50 to +3000 V DC

5.000 mA DC

Filter 1

50 to +5000 V dc

3.000 mA DC

Filter 2

Voltage set accuracy

2

: ±(0.01% of setting + 2.5 V)

Voltage display accuracy: voltage set accuracy ±1 V, typical (±2 V, maximum)

Voltage resolution: 1 V (set and display)

Voltage limit range: 0 to 100% full scale

Voltage regulation

3

:

Line: 0.001% for ±10% line voltage change

Load: 0.005% for 100% load change, typical

Output ripple (10 kHz – 100 kHz)

4

:

0.002% of full scale, V RMS, maximum - No filter

1.0 mV RMS @ 1 kV – Filter 1 or Filter 2

2.0 mV RMS @ 3 kV – Filter 1 or Filter 2

3.0 mV RMS @ 5 kV – Filter 2

Rise time (full load)

5, 6

:

1.5 seconds for 0 V setting to within 1 V of 5000 V DC – No filter

3 seconds for 0 V setting to within 1 V of 3000 V DC – Filter 1

4 seconds for 0 V setting to within 1 V of 5000 V DC – Filter 2

1

The output voltage can be programmed to a voltage lower than 50 V, however, performance below 50 V is not specified.

2

Add ±2.5 V DC when Filter 1 or Filter 2 is enabled.

3

Regulation specifications apply for greater than 25 V DC (with full load), or 50 V DC (with no load). Below these values, the unit may not

regulate correctly.

4

Peak-to-peak values are within five times the RMS value.

5

Current limit set to 105% of full scale.

6

Under resistive load.