Section 4 MAINTENANCE

4.1 Lamp Replacement

The procedure below explains lamp installation and electrical connections. Use a
small screwdriver to remove and install the lamp leads in the terminal block. The
instrument requires calibration after lamp replacement.

1. Orient the instrument so it is upside down and the top faces away from you.

Remove the battery cover and at least one battery.

2. Remove the lamp assembly by grasping the tab on the left side of the assembly

and remove the old lamp leads.

3. Insert the new lamp at the same way.

4. Replace the batteries and battery cover.

5. Insert the 800 FNU formazin standard into the sample cell. Press and hold

READ

.

Then press

I/O

. Release the

READ

key after the software version number disappears

from the display.

6. Adjust the scattered light amplifier output by inserting a small flat-blade

screwdriver into the trimpot hole (located on bottom). Adjust the display to read
2.5 ± 0.3 volts.

7. Press

I/O

to exit gain adjust mode.

8. Perform a formazin calibration according to Calibrating the Turbidimeter on

page 21.