Luminex 100 User Manual Version 1.7 Addendum

x

MAP Technology

6

PN 89-00002-00-143 Rev. B

Be sure the plate you are using is completely flat – warped plates can
lead to incorrect probe height adjustment.

In a 96-well microtiter plate where overall height is no more than
0.75 inches (19 mm), place the appropriate alignment tool in the
plate:

For a standard plate with flat-bottomed wells – stack two of the
larger (5.08 mm diameter) alignment discs together and place
them into the selected well.

For a filter-bottom plate – stack three of the larger (5.08 mm
diameter) alignment discs together and place them into the
selected well.

For a half-volume plate with flat-bottomed wells – stack two of
the smaller (3.35 mm diameter) alignment discs together and
place them into the selected well.

For a round-bottom (U-bottom) plate – stack two of the smaller
(3.35 mm diameter) alignment discs together and place them into
the selected well.

For a plate with conical wells – place one alignment sphere into
the selected well.

Sheath Fluid

USE OF NON-LUMINEX APPROVED SHEATH FLUID
SHALL CONSTITUTE “IMPROPER USE” AND MAY VOID
WARRANTY RIGHTS PROVIDED BY LUMINEX AND/OR
ITS AUTHORIZED PARTNER.