Luminex 100 User Manual Version 1.7 User Manual

Page 48

Advertising
background image

Luminex 100 User Manual Version 1.7

x

MAP Technology

5 - 8

PN 89-00002-00-063 Rev. A

3. Check the box in front of the report type you want. You can

check both boxes to see both kinds of calibration report.

4. Click OK to print the calibration report, or Preview to see the

report on your screen. Figure 5-10 shows an example of a
calibration log. Figure 5-11 shows an example of calibration
detail report.

Figure 5-10. Calibration Log

The calibration log shows gain values associated with each
calibration. A gain value is a representation of the voltage. Use this
log to observe the gain values. If you see a spike in voltage (20 V or
more) in the gain value for a channel from one calibration to the
next, the difference may indicate a problem with the instrument. It is
normal to see a value of -1 for RP1 gain for classification
calibrations. You also see gain values of -1 listed for DD, CL1, and
CL2 for reporter calibrations.