Run an entire plate of samples – Luminex 100 User Manual Version 1.7 User Manual

Page 75

Advertising
background image

x

MAP Technology

Data Acquisition

PN 89-00002-00-063 Rev. A

6 - 25

5. In the results table of the new session, highlight the sample

where you want to begin collecting data.

6. Click Start. You can see the progress of data collection on the

histogram and the bar graph. Make sure that the Single box
remains unchecked for continued sample acquisition.

7. After the analyzer finishes collecting data, select Save on the

File menu. The analyzer saves your data in the session folder
you selected when you created the session.

Run an Entire Plate of
Samples

‹

To run an entire plate:

1. Create the session for this run.

2. Click Eject from the toolbar to eject the plate holder.

3. Place a 96-well microtiter plate on the plate holder.

4. Click Retract from the toolbar to retract the plate holder.

5. Click Settings and make sure the number of samples is set to 96,

then click OK.

Figure 6-30. Settings Dialog Box

6. On the results table in the acquisition screen, highlight the row

for the first sample.

7. Click Start on the main control screen. The Luminex 100 auto-

matically reads the entire plate of samples.

Note:

If the Single box is checked, the software pauses between

samples. Click OK to continue acquiring sample.

8. After all samples have been read, click Eject from the toolbar to

eject the plate holder.

Note:

Ensure that the sample

needle height is adjusted to the
96-well plate in use. See page 7-9