Run a range of samples – Luminex 100 User Manual Version 1.7 User Manual

Page 76

Advertising
background image

Luminex 100 User Manual Version 1.7

x

MAP Technology

6 - 26

PN 89-00002-00-063 Rev. A

9. Remove the microtiter plate from the plate holder.

10. Click Retract from the toolbar to retract the plate holder.

Run a Range of Samples

You can run a specific range of samples, as demonstrated in the
highlighted samples shown below.

Figure 6-31. Sample Plate with Row of Samples

‹

To run a range of samples on a plate, follow these steps:

1. Create the session for this run.

2. Click Eject from the toolbar to eject the plate holder.

3. Place a 96-well microtiter plate on the plate holder.

4. Click Retract from the toolbar to retract the plate holder.

5. Click Settings and set the number of samples to the number of

the last sample that you want to read, then click OK.

To find the number of the last sample, start counting with well
A1, and count down an entire column. Then move to the next
column. For the example shown, you would set the number of
samples to 15.

6. On the results table part of the acquisition screen, select the sam-

ple where you want to begin collecting sample. For the example
shown, you would highlight well D2.

7. Click Start on the main control screen.

The Luminex 100 analyzer automatically reads the range of sam-
ples beginning with the selected sample.

8. After the samples have been read, click Eject from the toolbar to

eject the plate holder.

A
B
C
D
E
F
G
H

1 2 3 4 5 6 7 8 9 10 11 12

Advertising