Luminex 100 User Manual Version 1.7 User Manual

Page 85

Advertising
background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-063 Rev. A

7 - 5

2. On the back of the Luminex 100 analyzer, remove the thumb-

screw at the top of the air intake filter access door.

3. Pull the access door off of the Luminex 100 analyzer.

4. Pull the filter out from the unit, about three or four inches.

5. Grasp the tubing. Remove the filter with one hand, and hold the

tubing with the other hand.

6. Connect a new filter to the tubing.

7. Put the filter inside the panel, then reattach the panel door.

Figure 7-2. Rear of Luminex 100 Analyzer, Air Filter Exposed

‹

To replace the Luminex XYP Air Intake Filter:

1. Disconnect the Luminex XYP from ac power by turning off the

power switch on the rear of the instrument, then unplug the
power cord from the wall source.

2. On the back of the instrument, to the left side, gently remove the

screen from the filter. Do not remove the screws.

Figure 7-3. Back of Luminex XYP, Removing Air Filter