Change fuses – Luminex 100 User Manual Version 1.7 User Manual

Page 88

Advertising
background image

Luminex 100 User Manual Version 1.7

x

MAP Technology

7 - 8

PN 89-00002-00-063 Rev. A

Change the Sheath

Filter and

Quick-disconnect

O-rings

‹

Change the sheath quick-disconnect o-rings as necessary.

1. Turn off the Luminex 100 analyzer at the power switch.

2. Remove the quick-disconnect fitting that has the faulty o-ring.

3. Use a small flat-head screwdriver to remove the worn out o-ring.

4. Place the new o-ring over the tip of a pair of closed needle-nose

pliers.

5. Gently separate the tips of the pliers to stretch the o-ring just

enough to fit it over the top of the quick-disconnect.

6. Slide the o-ring over the quick disconnect, then remove the

pliers.

Change Fuses

‹

To replace fuses on the Luminex 100 analyzer or the Luminex

XYP instrument.

1. Turn off the power switch on both the Luminex 100 analyzer and

the Luminex XYP.

2. Unplug the power cord from the analyzer or instrument,

depending on which fuses you are changing.

3. Use a small, flathead screwdriver to open the power module door.

See Figure 7-6.

4. Remove the red cartridge. Check both fuses for damage.

5. Replace damaged fuses with the type specified on the sticker to

the right of the power input module.

6. Push the cartridge back into the power module, then push the

power module door back into place.

DANGER!

To avoid serious injury

or death by electric shock, you
must turn off the system and
unplug it from the wall outlet.

Advertising