Fluid leaks – Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 106

Advertising
background image

Luminex 200 User Manual for LDS Version 1.7

x

MAP Technology

8 - 4

PN 89-00002-00-150 Rev. A

Fluid Leaks

Fluid leaks can result in poor pressurization and failed sample
acquisition.

The compressor does not
engage.

Run a Prime command. If you
do not hear the compressor turn
on, call Technical Support.

The Cheminert Fitting is loose.

Ensure that the fitting connects
tightly above the sample probe,
below the blue light.

Fluid leaks in the system.

See Fluid Leak on page 8-4.

The sheath bottle has an air
leak.

Disconnect the sheath and waste
bottle connections from the
analyzer. Run a Prime
command. If pressure builds,
remove and retighten the sheath
fluid bottle cap, then reconnect
fluid lines to the analyzer. If
pressurization fails again,
replace the sheath bottle.

Pressure too High

The sheath bottle is overfilled.

Ensure that the sheath bottle is
not filled above the fill line.

Symptom

Possible Problem

Solution

Pressure too low

The sample probe is clogged.

Clean the sample probe. See
page 7-4.

The syringe seal leaks.

Replace the syringe seal. See
page 7-6.

The syringe valve leaks.

Hand-tighten the syringe
connection (silver knob) on the
syringe valve. Run a Prime
command. If leaks continue, call
Technical Support.

Large amount of fluid pooled
around instrument

Fittings or fluid lines are
damaged.

Contact Technical Support.

Fluid coming from sample
probe

The sample probe is clogged.

Clean the sample probe. See
page 7-4.