Sample probe – Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 107

Advertising
background image

x

MAP Technology

Troubleshooting

PN 89-00002-00-150 Rev. A

8 - 5

Sample Probe

Problems with the sample probe can lead to fluid leaks and
pressurization problems, as well as inhibit sample acquisition.

The sample 3-way valve is
faulty.

Contact Technical Support.

Fluid leaking from front of
analyzer

The syringe seal leaks.

Replace the syringe seal.

The syringe valve leaks.

Hand-tighten the syringe
connection (silver knob) to the
white syringe valve. Run a
Prime. If leaks continue, contact
Technical Support.

Symptom

Possible Problem

Solution

Sample probe leak

The sample probe is clogged.

Clean the sample probe. See
page 7-4.

Sample arm stuck in the up
position

The system isn’t properly
pressurized.

If a PRESSURIZATION error
message displays, see
Pressurization on page 8-3.