Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 108

Advertising
background image

Luminex 200 User Manual for LDS Version 1.7

x

MAP Technology

8 - 6

PN 89-00002-00-150 Rev. A

Sample arm stuck in the down
position

The sample probe height is too
low, or the path to the well is
blocked.

DO NOT turn off Luminex XYP
instrument.

1. Remove the blue light housing

on the Luminex analyzer.

2. Unscrew the tubing connector

connecting the sample tube to
the top of the sample arm.
The system monitor changes
from “Busy” to “Running”
and sample acquisition
continues. Click Cancel to
make adjustments before
continuing with the rest of the
samples.

3. If sample arm still does not

raise, save all the data that has
been collected to this point.
Turn off the analyzer, but
LEAVE THE XYP instrument
turned on.

4. Exit Data Collector.

5. Turn the analyzer back on and

restart the software.

6. Replace the sample tubing

and blue light. Adjust the
sample arm.

7. Run a calibration with water

to reset the sample probe
optical switch.

Sample arm does not go down
smoothly
.

The 96-well plate is incorrectly
seated in the XYP.

Adjust the 96-well plate.

The 96-well plate is warped.

Inspect the 96-well plate. Replace
it if it is warped.

The sample arm is misaligned.

Readjust the sample arm
horizontal alignment.

The analyzer is misaligned with
the XYP instrument.

Reposition the Luminex analyzer
on top of the Luminex XYP
instrument so that the holes
match up correctly.