MAP Technology

System Overview

PN 89-00002-00-150 Rev. A

3 - 11

Figure 3-7 Left Side of Analyzer - Air and Fluid Connectors

Sheath fluid container (not shown): Holds sheath fluid. For
proper operation, place the Luminex SD system at the same level as
the base of the Luminex XYP instrument. Do not put it on top of the
Luminex 200 analyzer. If you are not using the SD system, sheath
fluid levels must be monitored manually. Check the sheath fluid level
before starting a run or procedure.

Waste fluid container (not shown): The waste fluid container
receives waste from the system. The waste container should not be
placed on top of the instrument. Ensure that the waste tubing is not
elevated above the level of the Luminex 200 analyzer at any point.
Contact Luminex Technical Support before you relocate the waste
fluid container. To maintain a stable flow rate, do not move the waste
line or container during system operation.

Sheath Fluid
Connector (Blue)

Waste
Connector
(Orange)

Air Connector
(Green)

Warning:

If biological samples have been tested with the

system, use your standard laboratory safety practices.

Caution:

Waste levels must be manually monitored. Do not

allow the waste container to overflow

.