Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 40

Advertising
background image

Luminex 200 User Manual for LDS Version 1.7

x

MAP Technology

4 - 2

PN 89-00002-00-150 Rev. A

5. Place the Luminex analyzer on top of the Luminex XYP. Do not

plug it into the power outlet.

6. Without removing the Luminex analyzer alignment plate from

the black knob, shift the analyzer so it is centered on the
Luminex XYP.

7. Replace the short sample probe with the long sample probe as

follows: Unsnap the light housing above the probe. Loosen the
Cheminert fitting that holds the sample probe in place. Gently
grasp the probe and push up. Remove the sample probe out of
the top of the sample arm. Push the long sample probe into the
top of the sample arm, and tighten the Cheminert fitting.

8. Adjust the position of the Luminex analyzer with the Luminex

XYP.

9. Attach the waste container.

10. Fill and attach the sheath fluid container. Fill the container with

sheath fluid to just below the air intake.

‹

To set up the PC:

1. Place the PC to the right of the Luminex analyzer and Luminex

XYP. Do not place the PC on top of the analyzer.

2. Place the monitor on top of the PC and connect the cable

between the monitor and the PC.

3. Attach a power cable to the power input modules of the PC and

monitor. Do not plug the power cable into the power outlet.

4. Connect the mouse and keyboard to the PC.

5. Connect one of the communications cables between the Luminex

analyzer communications port and COM Port 1 on the PC.

6. Connect the other communications cable between the XYP

instrument communications port and the COM port 2 on the PC.

7. Plug all instrument power cords into an approved power source.