MAP Technology

System Startup

PN 89-00002-00-150 Rev. A

5 - 3

‹

To add a new lot of xMAP classification calibration

microspheres:

1. Click Calibrate. The Start Calibration dialog box opens.

Figure 5-2. Start Calibration Dialog Box

2. Click New in the Classification Channel box. The Add New

Classification Calibrator dialog box opens.

Figure 5-3. Add New Classification Calibrator Dialog Box

3. Enter the information into the dialog box, using the values on the

Certificate of Quality that came with your xMAP calibration
microspheres. Do not enter a value for CL3, but make sure to
click within the CL3 box and press Enter to save the other
values entered.

4. Click OK. You can now use this classification channel xMAP

calibration microsphere lot to calibrate your instrument.