MAP Technology

Data Acquisition

PN 89-00002-00-150 Rev. A

6 - 23

The message at the top reflects the current status of the Luminex
analyzer.

Ready: Indicates that the Luminex analyzer is ready to collect data.

Pressure: Pressure normally displays during a routine pressure
check at the beginning of each run. However, if the pressure message
remains for more than 1.5 minutes, a lack of pressure exists in the
system. Check for the following problems:

Low sheath fluid

Loose hose connections

Loose cap on sheath fluid bottle.

Leak in sheath fluid container, tubing, or bubble trap

Not Ready: Normally indicates that the analyzer is initializing.

Busy: Indicates an intermediate state between Standby and Ready.

Standby: Shows that the flow of sheath fluid stopped; the system is
idle.

Bad Link: The communication between the analyzer and the
computer is not working. Check that:

The analyzer is powered on.

The cables between the analyzer communications port and
computer are properly installed.

If the situation has been corrected, restart the Luminex software
application to reinitialize this link.

Warm Up: Indicates a 30-minute warmup of the analyzer’s optics.
While powered, the optics remain active for four hours after each
acquisition.

Running: Indicates that the Luminex analyzer is collecting data.

Sample empty: Indicates an empty sample syringe. The system
automatically cancels data collection and adds a comment in the
notes section.

Connect: Indicates that the communication between the analyzer
and the computer was broken, then re-established. You must restart
the Luminex software application to reinitialize communication
links.

Sheath fluid: Indicates empty sheath fluid. Fill the sheath fluid
container before continuing.