Run a single sample, Interrupt a run – Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 85

Advertising
background image

x

MAP Technology

Data Acquisition

PN 89-00002-00-150 Rev. A

6 - 27

9. Remove the microtiter plate from the plate holder.

10. Click Retract from the toolbar to retract the plate holder.

Run a Single Sample

‹

To run a single sample:

1. Click Eject from the toolbar to eject the plate holder.

2. Place a 96-well microtiter plate on the plate holder.

3. Click Retract from the toolbar to retract the plate holder.

4. On the results table part of the acquisition screen, select the sam-

ple that you want to read.

5. Check the Single checkbox below the start button on the main

control screen.

6. Click Start on the main control screen. The Luminex 200 reads

just the selected sample.

7. Click Eject from the toolbar to eject the plate holder.

8. Remove the microtiter plate from the plate holder.

9. Click Retract from the toolbar to retract the plate holder.

Interrupt a Run

During Sample

Acquisition

‹

To interrupt a run while the system acquires a sample:

1. Click Cancel. Sampling stops after you click Cancel, and a win-

dow pops up asking if you want to acquire the next sample. If
you click Yes, sampling proceeds with the next sample. Click No
to stop the session completely.

2. To continue acquiring sample, highlight the sample with which

you want to resume sampling with and click Start.

Between Samples

The Single checkbox becomes accessible for several seconds
between samples. Click the Single checkbox. The software then
pauses between acquiring the next sample run.

Click OK to continue acquiring the next sample.