Clean luminex analyzer ventilation filter, Change sheath filter – Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 95

Advertising
background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-150 Rev. A

7 - 7

1. Syringe Seal 2. Syringe

Figure 7-3. Front of Luminex 200 Analyzer, Access Door Opened

Clean Luminex

Analyzer Ventilation

Filter

‹

Check the Luminex analyzer ventilation filter every 6 months.

Clean the filter only when soiled.

1. Turn off the Luminex analyzer at the power switch.

2. Unplug the power cord from the wall source.

3. On the bottom of the unit on the left side, push the filter out

toward the right side.

4. Grasp the filter and pull it out from underneath the Luminex

analyzer.

5. If the filter is dirty, clean it with a vacuum cleaner or in running,

distilled water. Stand it upright to air-dry.

6. Reinstall the filter with the arrows facing up.

Change Sheath

Filter

‹

Change the sheath filter once a year.

1. Disconnect the sheath fluid bottle

2. Squeeze the metal quick release connections at both ends of the

filter assembly, then pull the tubing away from the filter
assembly.