Luminex xyp instrument firmware upgrade, Turn on power to the xyp instrument, Close any open applications – Luminex IS 2.1_2.2 to xPONENT 3.1 Rev 2 Upgrade Instructions User Manual

Page 14

Advertising
background image

Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use

10

LX200 Sample Probe

5. The Download In Progress dialog box opens displaying the

progress. Read the information presented on the dialog box.
When the download is finished the Status text box displays
Download Complete. Click OK.

6. Turn off the power to the analyzer.

7. Replace the serial-to-serial cable from the analyzer to the

computer with the new USB cable supplied in the upgrade kit by
plugging the serial end into P1 on the analyzer, and the USB end
into an available USB port on the computer.

8. Connect the short CANBUS cable (gray) to P1 on the back of the

SD and P2 on the back of the LX100 or LX200 analyzer.

9. Turn the power on to the Luminex analyzer.

10. Turn on the power to the SD.

11. An entry in the Detector Device text box will display the new

firmware version. For the Luminex 100 analyzer, the firmware is
updated to 2.3.9. For the Luminex 200 analyzer, the firmware is
updated to 2.6.1.

Luminex XYP Instrument

Firmware Upgrade

To upgrade the Luminex XYP instrument firmware:

1. Turn on power to the XYP instrument.
2. Close any open applications.
3. On the PC desktop click Start > Programs > Luminex > LXR >

LxFirmware. The LxFirmware Update Tool dialog box opens.
See "LxFirmware Update Tool Dialog Box" on page 8. The refer-
enced graphic is just an example; the dialog box on your system
may show different upgrade versions.

4. Select the Movement option, if necessary. Wait until the top line

of the Luminex XYP/SD Instrument Firmware Version text box