Adding an external analysis program, Editing an analysis program, Removing an analysis program – Luminex xPONENT 4.2 for FLEXMAP User Manual

Page 126: Arranging main navigation buttons

Advertising
background image

System Initialization Procedure - Displays the available selections to use as the default

system initialization procedure. After selecting an option, click Save at the bottom right
portion of the window to save the selection as the default procedure.

Allow running calibration or verification if the instrument is not warmed up - You can

enable or diable this feature by selecting the check box.

Calibration expiration days - Allows to set how often the system will need to be re-

calibrated.

Adding an External Analysis Program

To add an external analysis program:

1. If the program is on an external media such as a CD or flash drive, insert the media.

2. Click Add New to open the New External Analysis Program dialog box.

3. Type a name for the external analysis program.

4. Click Browse to navigate to the .exe file for that program. Double-click the file name.

5. Type the command line parameter for the parameters you want xPONENT to use with

the external analysis program. If the information is supplied with the external analysis
program documentation, use that information. Otherwise, you can type the following
parameters built into xPONENT, in any order:

#c - Output.csv, full file path.

#p - Protocol name.

#b - Batch name.

#u - Logged in user name.

To keep the default command line settings, leave Command Line Parameters blank.

Editing an Analysis Program

1. In the Installed Analysis Programs list, click the program you want to edit.

2. Click Edit. The Edit External Analysis Program dialog box opens.

3. Edit the Name, Path, or Command Line Parameters, or make this the default analysis

program if there are two or more programs installed. The default analysis program name
displays in bold text.

Removing an Analysis Program

To remove an analysis program from the Installed Analysis Programs list:

1. In the Installed Analysis Programs list, select the program you want to uninstall.

2. Click Remove. To prevent the external analysis program from starting automatically,

select Disable automatic launching of External Analysis when batches complete for
all protocols
.

Arranging Main Navigation Buttons

Use this section to arrange the main pages at the top of the xPONENT

®

screen.

xPONENT

®

for FLEXMAP 3D

®

112

Advertising