NOTE: The Home page, and in some instances the Admin page, cannot be

moved.

To arrange the main navigation buttons:

1. Select or clear the check boxes by each page name to hide or display the page.

2. Click a page name and use the up and down arrows to change the order in which the

pages display, from left to right.

3. Click Save.

4. Click Default if you want to restore the main navigation.

Maintenance Options

This section allows you to define the System Initialization Routine.

Run one of the System Initialization procedures as part of your daily startup routine.

NOTE: See "Daily Activities" on the Home page for daily maintenance

routines.

Luminex recommends that you verify daily and calibrate weekly. You should also verify and
calibrate if any of the following occurs:

• The delta calibration temperature exceeds ± 5° C.

• You move the instrument.

• You experience sample acquisition problems.

• You perform maintenance on the instrument, for example, replacing a hardware

component.

Available system initialization procedures are as follows:

• Fluidics preparation, calibration, performance verification

• Fluidics preparation, performance verification

• Fluidics preparation

Defining the System Initialization

NOTE: Luminex

®

recommends daily performance verification and weekly

calibration of the FLEXMAP 3D

®

system. You can set up the system

initialization routine to include calibration and verification on the
Admin page, System Setup tab, Maintenance Options section.

1. Open the Admin page.

Admin Page

113