Luminex IS Version 2.3 User Manual

Page 63

Advertising
background image

x

MAP Technology

Luminex 2.3 Software

PN 89-00002-00-110 Rev. B

53

batch. If you reprocess a batch using changed parameters, the system
may obtain different results. When you replay a batch it labels
unknown samples as Pa1, Pa2, Pa3, and so on. If you replay a batch
containing replicates, replicate averaging will not be calculated in
data analysis.

A number of variables can affect the final test results. You may also
change the standards or controls processed with the batch or multi-
batch. These variables may effect your test results:

minimum number of events for acquisition

formula used to analyze the MFI values

standards or controls validation or invalidation

type of analysis (qualitative, quantitative, acquisition only, or
maintenance)

‹

To reprocess samples using Replay Batch:

1. On the Acq. Detail tab, click Replay Batch. The Browse for

Folder dialog box opens displaying the My Batches folder.

2. Select the desired batch under the My Batches folder and click

OK.

3. The Open Template dialog box opens. Click on the desired

template and click Select.

4. The Run Batch tab becomes the active tab. You can monitor the

commands as they process. Click on the Acq. Detail tab and
monitor the data, histogram, and dot plot.

After replaying a batch, you can analyze the data.

‹

To analyze Replay Batch data:

1. Click Start Analysis.

2. In the Open Batch dialog box, select the batch you want to

analyze, then click Select. The most recent Replay Batch is the
last or has the highest ID number. See Figure 29.

Note:

The Open Batch

dialog box does not list or
show the templates
associated to the batches.

Advertising