Invalidate Standards and Controls

Your xPONENT system administrator must give you privileges to invalidate standards if
you are using the Secure xPONENT package. You can invalidate or remove a standard if
doing so improves the curve fit.

CAUTION: Observe caution when invalidating standards. Doing so greatly

affects the curve fit and subsequent sample results.

For further instruction on assay standard curves and the appropriateness of invalidating or
removing standards, contact the assay kit manufacturer.

It is possible to invalidate or remove a control in data analysis. However, Luminex does
not recommend invalidating controls. For information on assay controls and guidelines
regarding accepting or rejecting control values, contact the assay kit manufacturer.

Your system administrator must give you privileges to invalidate standards.

1. Open the Results page, then open the Saved Batches tab.
2. Open the Saved Batches tab.
3. Click the batch name, then click Open. The Results tab opens.
4. Click the square area next to left of the standard you wish to invalidate, then click

Invalidate. The whole row turns red.

View Batch Settings

1. Open the Results page, then open the Saved Batches tab.
2. Click Saved Batches, then click the batch for which you want to view details.
3. Click Open, then click the Settings tab.
4. Click the left and right Page arrows to view the pages of the batch settings report.
5. Click Save to open the Save As dialog box. Navigate to the location you wish to save

the batch settings report, and click Save.

View Batch Logs

1. Open the Results page, then open the Saved Batches tab.
2. Click Saved Batches, then click the batch for which you want to view details.
3. Click Open. The Results tab opens.
4. Click Log to open the Log tab.

View Sample Details

1. Open the Results page., then open the Saved Batches tab.
2. Click Saved Batches, then click the batch for which you want to view details.
3. Click Open, then click Sample Details. The Sample Details tab opens. If you are

using an LIS licensed package of the software, click Transmit to transmit sample
details to the LIS database. You can transmit either a single analyte per sample or the
entire sample.