System monitor – Luminex xPONENT 4.0 SP1 User Manual

Page 22

Advertising
background image

Luminex xPONENT 4.0 Software Manual

10

Export - Opens a File Dialog dialog box. Use the Browse button to select a location,
file name, and file type (either a text.CSV file) for the export. This exports all data from

the right-clicked item.

Cut - Cuts the selected data.

Copy All - Copies all data.

Copy - Copies only the selected data.

Paste - Pastes previously copied text or data into the box.

Delete - Erases text or data from the selection.

The histogram, dot plot, and 3D View within the Results page have specialty right-click
menus. See those sections for more information.

System Monitor

The System Monitor remains at the bottom of all xPONENT

®

windows. It displays the

physical state of the Luminex analyzer. The values displayed are reported directly from the
Luminex system.

FIGURE 3.

System Monitor

System Status. This button has two functions: When clicked, it opens the system log (see
“System Status Tab” on page 79)
. It also displays the current status of the system. If there
are no warnings or errors, the System Status button is green with a check mark. If there is
a warning, error, or other important user-notification, the button turns yellow with an
exclamation point.

Command. Displays the command currently running on the system.

NOTE:

A yellow Check Calibration button is visible in the command
display if the calibration or verification command has failed, when
any verification is not current, or when the calibration or
verification was performed prior to the calibration expiration
setting.

1 System Status button

2 Command display

3 Progress bar and buttons

4 Delta Cal Temp status button

5 Sheath Pressure status

6 XY status

7 Laser status

8 Region Events status

9 Total events status

1

2

3

4 5

6

7

9

8

Advertising